我们有一个E4432B,它的波形存储器有问题。
看起来内存中有多个波形,您无法始终选择合适的波形。 它有时会播放错误的波形,甚至在某些情况下,两个波形的组合! 只要您一次只加载一个波形,或者只是将它用作CW发生器,发生器仍然可以使用。当发生此问题时,固件处于修订版本:B.03.86我尝试将固件降级到B.03.84,但是 问题仍然存在。有谁知道我能解决的这个问题是否可以通过处理wavefore序列SRAM的服务说明E4432B-11解决?感谢您的时间。 以上来自于谷歌翻译 以下为原文 We have an E4432B that has a problem with its waveform memory. It seems like if there is more than one waveform in memory, you can’t always select the proper waveform. It will sometimes play out the wrong waveform or even in some cases, a combination of two waveforms! The generator is still somewhat usable as long as you only load one waveform at a time or else just use it as a CW generator. When this problem occurred firmware was at revision: B.03.86 I tried downgrading the firmware to B.03.84, but the problem is still present. Does anyone know if this issue I am having could be resolved with Service note E4432B-11 which deal with wavefore sequence SRAM? Thanks for your time. |
|
相关推荐
2个回答
|
|
嗨 - 这听起来像是服务说明E4432B-11中提到的双任意板上的缺陷,其中回放期间的波形失真。
您很可能需要在当地的安捷伦服务中心维修设备。 以上来自于谷歌翻译 以下为原文 Hi - This sounds like a defect on the dual Arbitrary board mentioned in the Service Note E4432B-11 where the waveform during playback became distorted. Most likely you will need to have the unit repaired at your local Agilent service center. Regards - |
|
|
|
|
|
|
|
只有小组成员才能发言,加入小组>>
855 浏览 0 评论
2188 浏览 1 评论
1971 浏览 1 评论
1834 浏览 5 评论
2708 浏览 3 评论
680浏览 1评论
关于Keysight x1149 Boundary Scan Analyzer
478浏览 0评论
N5230C用“CALC:MARK:BWID?”获取Bwid,Cent,Q,Loss失败,请问大佬们怎么解决呀
574浏览 0评论
2366浏览 0评论
1420浏览 0评论
小黑屋| 手机版| Archiver| 电子发烧友 ( 湘ICP备2023018690号 )
GMT+8, 2024-8-18 05:39 , Processed in 1.222382 second(s), Total 48, Slave 42 queries .
Powered by 电子发烧友网
© 2015 www.ws-dc.com
关注我们的微信
下载发烧友APP
电子发烧友观察
版权所有 © 湖南华秋数字科技有限公司
电子发烧友 (电路图) 湘公网安备 43011202000918 号 电信与信息服务业务经营许可证:合字B2-20210191 工商网监 湘ICP备2023018690号